Purpose: To characterize the artifacts occurring in helical 4D CT imaging and propose a method to identify the artifacts automatically. Method and Material:_Given two adjacent stacks obtained from same respiration phase, we seek to determine if there are artifacts between them. The proposed method uses a “bridge” stack strategy to connect two stacks. Using normalized cross correlation convolution (NCCC), two stacks are mapped to the “bridge” stack and the best matching positions can be located. Using the position information, the method can determine if there is an artifact or not between the two stacks. Furthermore, by combining the matching positions with NCCC values, the performance can be improved. To validate the method, sensitivity and specificity are used as the metrics to evaluate the method with 500 hundred stacks on five patients which are labeled by three physicists independently. Results: Three types of artifact sources are categorized: low frequency respiration, high amplitude period and low amplitude period. Low frequency respiration can cause blurring. Both amplitude period and low amplitude period can cause anatomy gap and overlap occur depend on different anatomy motion directions. The artifacts identification results show that four of the five patients' data are validated using the proposed method. The average sensitivity is about 83% and the average specificity 84%. Furthermore, the results show that the method improves sensitivity to 87.8% and specificity to 89% when combining matching positions with NCCC values. Conclusion: This study shows that the spatial artifacts during 4D CT imaging are characterized and can be located by the proposed method. It is valuable to quantify the quality of 4D CT image and reduce the artifacts directly from the reconstructed images.
Details
- Title: Subtitle
- SU‐GG‐I‐128: Characterization and Identification of Artifacts for Helical 4D CT Image
- Creators
- D Han - 3University of Iowa, Department of Electrical and Computer Engineering, Iowa City, IAJ Bayouth - 3University of Iowa, Department of Electrical and Computer Engineering, Iowa City, IAS Bhatia - 3University of Iowa, Department of Electrical and Computer Engineering, Iowa City, IAM Sonka - 3University of Iowa, Department of Electrical and Computer Engineering, Iowa City, IAX Wu - 3University of Iowa, Department of Electrical and Computer Engineering, Iowa City, IA
- Resource Type
- Conference proceeding
- Publication Details
- Medical physics (Lancaster), Vol.37(6), pp.3131-3131
- DOI
- 10.1118/1.3468162
- ISSN
- 0094-2405
- eISSN
- 2473-4209
- Number of pages
- 1
- Language
- English
- Date published
- 06/2010
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Electrical and Computer Engineering; Radiation Oncology; Fraternal Order of Eagles Diabetes Research Center; Injury Prevention Research Center; Ophthalmology and Visual Sciences
- Record Identifier
- 9984047653102771
Metrics
17 Record Views