Metal Ion Sensitivity
Encyclopedia entry

Metal Ion Sensitivity

William M. Mihalko, Catherine R. Olinger and Boyle C. Cheng
Handbook of Spine Technology, pp.459-471
Springer International Publishing
04/01/2021
DOI: 10.1007/978-3-319-44424-6_139

View Online

Abstract

Metal hypersensitivity to biomaterial alloys have been reported since the 1970s. While most reports have been in the total joint literature, in the last 10 years isolated spinal implant reactions have been reported. Much of this is because spine implants have been developed with bearing surfaces that may be a trigger for sensitizing patient from the local wear debris. Reaction to metal alloys and debris is a type IV hypersensitivity immunologic reaction in that it does not produce anaphylaxis. The adverse local tissue reactions (ALTR) around the implant can be substantial and lead to further surgery. The metal alloys used in spinal implants typically have an oxide passivation layer that can protect the body from these local reactions, but any type of fretting from modular connections of wear from a metal bearing can lead to exposure of the alloy below the passivation layer and be the trigger to the start of a reaction leading to ALTR. Knowing the frequency of these sensitivities in the general population can help surgeons identify hypersensitive patients and notify them of the possible risk.
Biomaterial Hypersensitivity Metal alloy Passivation layer