3D dosimetric validation of ultrasound-guided radiotherapy with a dynamically deformable abdominal phantom
Physica medica, Vol.84, pp.159-167
04/2021
DOI: 10.1016/j.ejmp.2021.04.007
PMCID: PMC8216233
PMID: 33901860
Abstract
The purpose of this study was to dosimetrically benchmark gel dosimetry measurements in a dynamically deformable abdominal phantom for intrafraction image guidance through a multi-dosimeter comparison. Once benchmarked, the study aimed to perform a proof-of-principle study for validation measurements of an ultrasound image-guided radiotherapy delivery system.
The phantom was dosimetrically benchmarked by delivering a liver VMAT plan and measuring the 3D dose distribution with DEFGEL dosimeters. Measured doses were compared to the treatment planning system and measurements acquired with radiochromic film and an ion chamber. The ultrasound image guidance validation was performed for a hands-free ultrasound transducer for the tracking of liver motion during treatment.
Gel dosimeters were compared to the TPS and film measurements, showing good qualitative dose distribution matches, low γ values through most of the high dose region, and average 3%/5 mm γ-analysis pass rates of 99.2%(0.8%) and 90.1%(0.8%), respectively. Gel dosimeter measurements matched ion chamber measurements within 3%. The image guidance validation study showed the measurement of the treatment delivery improvements due to the inclusion of the ultrasound image guidance system. Good qualitative matching of dose distributions and improvements of the γ-analysis results were observed for the ultrasound-gated dosimeter compared to the ungated dosimeter.
DEFGEL dosimeters in phantom showed good agreement with the planned dose and other dosimeters for dosimetric benchmarking. Ultrasound image guidance validation measurements showed good proof-of-principle of the utility of the phantom system as a method of validating ultrasound-based image guidance systems and potentially other image guidance methods.
Details
- Title: Subtitle
- 3D dosimetric validation of ultrasound-guided radiotherapy with a dynamically deformable abdominal phantom
- Creators
- Charles K Matrosic - University of Wisconsin–MadisonWesley Culberson - University of Wisconsin–MadisonAndrew Shepard - University of Wisconsin–MadisonSydney Jupitz - University of Wisconsin–MadisonBryan Bednarz - University of Wisconsin–Madison
- Resource Type
- Journal article
- Publication Details
- Physica medica, Vol.84, pp.159-167
- DOI
- 10.1016/j.ejmp.2021.04.007
- PMID
- 33901860
- PMCID
- PMC8216233
- NLM abbreviation
- Phys Med
- ISSN
- 1120-1797
- eISSN
- 1724-191X
- Grant note
- R01 CA190298 / NCI NIH HHS
- Language
- English
- Date published
- 04/2021
- Academic Unit
- Radiation Oncology
- Record Identifier
- 9984313071202771
Metrics
15 Record Views