A controlled statistical study to assess measurement variability as a function of test object position and configuration for automated surveillance in a multicenter longitudinal COPD study (SPIROMICS)
Medical physics (Lancaster), Vol.43(5), pp.2598-2610
05/2016
DOI: 10.1118/1.4947303
PMCID: PMC4851622
PMID: 27147369
Abstract
A test object (phantom) is an important tool to evaluate comparability and stability of CT scanners used in multicenter and longitudinal studies. However, there are many sources of error that can interfere with the test object-derived quantitative measurements. Here the authors investigated three major possible sources of operator error in the use of a test object employed to assess pulmonary density-related as well as airway-related metrics.
Two kinds of experiments were carried out to assess measurement variability caused by imperfect scanning status. The first one consisted of three experiments. A COPDGene test object was scanned using a dual source multidetector computed tomographic scanner (Siemens Somatom Flash) with the Subpopulations and Intermediate Outcome Measures in COPD Study (SPIROMICS) inspiration protocol (120 kV, 110 mAs, pitch = 1, slice thickness = 0.75 mm, slice spacing = 0.5 mm) to evaluate the effects of tilt angle, water bottle offset, and air bubble size. After analysis of these results, a guideline was reached in order to achieve more reliable results for this test object. Next the authors applied the above findings to 2272 test object scans collected over 4 years as part of the SPIROMICS study. The authors compared changes of the data consistency before and after excluding the scans that failed to pass the guideline.
This study established the following limits for the test object: tilt index ≤0.3, water bottle offset limits of [-6.6 mm, 7.4 mm], and no air bubble within the water bottle, where tilt index is a measure incorporating two tilt angles around x- and y-axis. With 95% confidence, the density measurement variation for all five interested materials in the test object (acrylic, water, lung, inside air, and outside air) resulting from all three error sources can be limited to ±0.9 HU (summed in quadrature), when all the requirements are satisfied. The authors applied these criteria to 2272 SPIROMICS scans and demonstrated a significant reduction in measurement variation associated with the test object.
Three operator errors were identified which significantly affected the usability of the acquired scan images of the test object used for monitoring scanner stability in a multicenter study. The authors' results demonstrated that at the time of test object scan receipt at a radiology core laboratory, quality control procedures should include an assessment of tilt index, water bottle offset, and air bubble size within the water bottle. Application of this methodology to 2272 SPIROMICS scans indicated that their findings were not limited to the scanner make and model used for the initial test but was generalizable to both Siemens and GE scanners which comprise the scanner types used within the SPIROMICS study.
Details
- Title: Subtitle
- A controlled statistical study to assess measurement variability as a function of test object position and configuration for automated surveillance in a multicenter longitudinal COPD study (SPIROMICS)
- Creators
- Junfeng Guo - Departments of Radiology and Biomedical Engineering, University of Iowa, Iowa City, Iowa 52242Chao Wang - Department of Statistics and Actuarial Science, University of Iowa, Iowa City, Iowa 52242Kung-Sik Chan - Department of Statistics and Actuarial Science, University of Iowa, Iowa City, Iowa 52242Dakai Jin - Department of Electrical and Computer Engineering, University of Iowa, Iowa City, Iowa 52242Punam K Saha - Department of Electrical and Computer Engineering, University of Iowa, Iowa City, Iowa 52242Jered P Sieren - Department of Radiology, University of Iowa, Iowa City, Iowa 52242R G Barr - Departments of Medicine and Epidemiology, Columbia University Medical Center, New York, New York 10032MeiLan K Han - Department of Medicine, Division of Pulmonary and Critical Care Medicine, University of Michigan, Ann Arbor, Michigan 48109Ella Kazerooni - Department of Radiology, University of Michigan, Ann Arbor, Michigan 48109Christopher B Cooper - Department of Medicine, University of California, Los Angeles, California 90095David Couper - Department of Biostatistics, University of North Carolina, Chapel Hill, North Carolina 27599John D Newell Jr - Departments of Radiology and Biomedical Engineering, University of Iowa, Iowa City, Iowa 52242Eric A Hoffman - Departments of Radiology, Medicine and Biomedical Engineering, University of Iowa, Iowa City, Iowa 52242
- Resource Type
- Journal article
- Publication Details
- Medical physics (Lancaster), Vol.43(5), pp.2598-2610
- DOI
- 10.1118/1.4947303
- PMID
- 27147369
- PMCID
- PMC4851622
- NLM abbreviation
- Med Phys
- ISSN
- 0094-2405
- eISSN
- 2473-4209
- Publisher
- United States
- Grant note
- HHSN268200900017C / NHLBI NIH HHS P30 ES005605 / NIEHS NIH HHS HHSN268200900019C / NHLBI NIH HHS HHSN268200900020C / NHLBI NIH HHS HHSN268200900015C / NHLBI NIH HHS HHSN268200900013C / NHLBI NIH HHS R01 HL112986 / NHLBI NIH HHS P30 CA086862 / NCI NIH HHS HHSN268200900016C / NHLBI NIH HHS HHSN268200900018C / NHLBI NIH HHS HHSN268200900014C / NHLBI NIH HHS S10 OD018526 / NIH HHS P30 DK054759 / NIDDK NIH HHS
- Language
- English
- Date published
- 05/2016
- Academic Unit
- Statistics and Actuarial Science; Roy J. Carver Department of Biomedical Engineering; Radiology; Electrical and Computer Engineering; Internal Medicine
- Record Identifier
- 9984051519202771
Metrics
38 Record Views