Logo image
A randomised trial comparing the i-gelTM with the LMA ClassicTM in children
Journal article   Peer reviewed

A randomised trial comparing the i-gelTM with the LMA ClassicTM in children

J. -R. Lee, M. -S. Kim, J. -T. Kim, H. -J. Byon, Y. -H. Park, H. -S. Kim and C. -S. Kim
Anaesthesia, Vol.67(6), pp.606-611
06/01/2012
DOI: 10.1111/j.1365-2044.2012.07072.x
PMID: 22352745

View Online

Abstract

We performed a prospective, randomised trial comparing the i-gelTM with the LMA ClassicTM in children undergoing general anaesthesia. Ninety-nine healthy patients were randomly assigned to either the i-gel or the LMA Classic. The outcomes measured were airway leak pressure, ease of insertion, time taken for insertion, fibreoptic examination and complications. Median (IQR [range]) time to successful device placement was shorter with the i-gel (17.0 (13.8-20.0 [10.0-20.0]) s) compared with the LMA Classic (21.0 (17.5-25.0 [15.0-70.0]) s, p = 0.002). There was no significant difference in oropharyngeal leak pressure between the two devices. A good fibreoptic view of the glottis was obtained in 74% of the i-gel group and in 43% of the LMA Classic group (p < 0.001). There were no significant complications. In conclusion, the i-gel provided a similar leak pressure, but a shorter insertion time and improved glottic view compared with the LMA Classic in children.
Anesthesiology Life Sciences & Biomedicine Science & Technology

Details

Metrics

4 Record Views
Logo image