Accuracy of linear measurement in the Galileos cone beam computed tomography under simulated clinical conditions
Dento-maxillo-facial radiology, Vol.40(5), pp.299-305
07/2011
DOI: 10.1259/dmfr/72117593
PMCID: PMC3520257
PMID: 21697155
Abstract
The aim of this study was to determine the geometric accuracy of cone beam CT (CBCT)-based linear measurements of bone height obtained with the Galileos CBCT (Sirona Dental Systems Inc., Bensheim, Hessen, Germany) in the presence of soft tissues.\nSix embalmed cadaver heads were imaged with the Galileos CBCT unit subsequent to placement of radiopaque fiduciary markers over the buccal and lingual cortical plates. Electronic linear measurements of bone height were obtained using the Sirona software. Physical measurements were obtained with digital calipers at the same location. This distance was compared on all six specimens bilaterally to determine accuracy of the image measurements.\nThe findings showed no statistically significant difference between the imaging and physical measurements (P > 0.05) as determined by a paired sample t-test. The intraclass correlation was used to measure the intrarater reliability of repeated measures and there was no statistically significant difference between measurements performed at the same location (P > 0.05).\nThe Galileos CBCT image-based linear measurement between anatomical structures within the mandible in the presence of soft tissues is sufficiently accurate for clinical use.
Details
- Title: Subtitle
- Accuracy of linear measurement in the Galileos cone beam computed tomography under simulated clinical conditions
- Creators
- R Ganguly - Oral and Maxillofacial Radiology, General Dentistry, Tufts University School of Dental Medicine, Boston, MA, USA. rumpa.ganguly@tufts.eduA RuprechtS VincentJ HellsteinS TimmonsF Qian
- Resource Type
- Journal article
- Publication Details
- Dento-maxillo-facial radiology, Vol.40(5), pp.299-305
- Publisher
- England
- DOI
- 10.1259/dmfr/72117593
- PMID
- 21697155
- PMCID
- PMC3520257
- ISSN
- 0250-832X
- eISSN
- 1476-542X
- Language
- English
- Date published
- 07/2011
- Academic Unit
- Preventive and Community Dentistry; Radiology; Oral Pathology, Radiology and Medicine; Anatomy and Cell Biology; Pathology; Dentistry Administration; Dental Research
- Record Identifier
- 9984066089602771
Metrics
36 Record Views