Journal article
An in vitro evaluation of the pressure generated during programmed intermittent epidural bolus injection at varying infusion delivery speeds
Journal of clinical anesthesia, Vol.34, pp.632-637
11/2016
DOI: 10.1016/j.jclinane.2016.06.017
PMID: 27687462
Abstract
Programmed intermittent bolus injection of epidural anesthetic solution results in decreased anesthetic consumption and better patient satisfaction compared with continuous infusion, presumably by better spread of the anesthetic solution in the epidural space. It is not known whether the delivery speed of the bolus injection influences analgesia outcomes. The objective of this in vitro study was to determine the pressure generated by a programmed intermittent bolus pump at 4 infusion delivery speeds through open-ended, single-orifice and closed-end, multiorifice epidural catheters.
In vitro observational study.
Not applicable.
Not applicable.
A CADD-Solis Pain Management System v3.0 with Programmed Intermittent Bolus Model 2110 was connected via a 3-way adapter to an epidural catheter and a digital pressure transducer. Pressures generated by delivery speeds of 100, 175, 300, and 400 mL/h of saline solution were tested with 4 epidural catheters (2 single orifice and 2 multiorifice). These runs were replicated on 5 pumps. Analysis of variance was used to compare the mean peak pressures of each delivery speed within each catheter group (single orifice and multiorifice).
Thirty runs at each delivery speed were performed with each type of catheter for a total of 240 experimental runs. Peak pressure increased with increasing delivery speeds in both catheter groups (P<.001). Peak pressures were higher with the multiorifice catheter compared with the single-orifice catheter at all delivery speeds (P<.001, for all).
Using a pump designed for programmed intermittent infusion boluses, the delivery speed of saline solution through epidural catheters was directly related to the peak pressures. Future work should evaluate whether differences in the delivery speed of anesthetic solution into the epidural space correlate with differences in the duration and quality of analgesia during programmed intermittent epidural bolus delivery.
•Higher epidural infusion delivery speeds are associated with higher peak pressures.•At all speeds, peak pressures are higher in multiorifice vs single-orifice catheters.•The clinical implications of these pressure differences need to be evaluated.
Details
- Title: Subtitle
- An in vitro evaluation of the pressure generated during programmed intermittent epidural bolus injection at varying infusion delivery speeds
- Creators
- Thomas T Klumpner - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611Elizabeth M.S Lange - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611Heena S Ahmed - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611Paul C Fitzgerald - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611Cynthia A Wong - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611Paloma Toledo - Department of Anesthesiology, Northwestern University Feinberg School of Medicine, 251 E Huron St, F5-704, Chicago, IL, USA 60611
- Resource Type
- Journal article
- Publication Details
- Journal of clinical anesthesia, Vol.34, pp.632-637
- Publisher
- Elsevier Inc
- DOI
- 10.1016/j.jclinane.2016.06.017
- PMID
- 27687462
- ISSN
- 0952-8180
- eISSN
- 1873-4529
- Grant note
- 69779 / Robert Wood Johnson Foundation Harold Amos Medical Faculty Development
- Language
- English
- Date published
- 11/2016
- Academic Unit
- Anesthesia
- Record Identifier
- 9984006353102771
Metrics
35 Record Views