An organ and effective dose study of XVI and OBI cone‐beam CT systems
Journal of applied clinical medical physics, Vol.11(2), pp.181-197
2010
DOI: 10.1120/jacmp.v11i2.3183
PMCID: PMC5719945
PMID: 20592702
Abstract
The main purpose of this work was to quantify patient organ doses from the two kilovoltage cone beam computed tomography (CBCT) systems currently available on medical linear accelerators, namely the X‐ray Volumetric Imager (XVI, Elekta Oncology Systems) and the On‐Board Imager (OBI, Varian Medical Systems). Organ dose measurements were performed using a fiber‐optic coupled (FOC) dosimetry system along with an adult male anthropomorphic phantom for three different clinically relevant scan sites: head, chest, and pelvis. The FOC dosimeter was previously characterized at diagnostic energies by Hyer et al. [Med Phys 2009;36(5):1711–16] and a total uncertainty of approximately 4% was found for in‐phantom dose measurements. All scans were performed using current manufacturer‐installed clinical protocols and appropriate bow‐tie filters. A comparison of image quality between these manufacturer‐installed protocols was also performed using a Catphan 440 image quality phantom. Results indicated that for the XVI, the dose to the lens of the eye (1.07 mGy) was highest in a head scan, thyroid dose (19.24 mGy) was highest in a chest scan, and gonad dose (29 mGy) was highest in a pelvis scan. For the OBI, brain dose (3.01 mGy) was highest in a head scan, breast dose (5.34 mGy) was highest in a chest scan, and gonad dose (34.61 mGy) was highest in a pelvis scan. Image quality measurements demonstrated that the OBI provided superior image quality for all protocols, with both better spatial resolution and low‐contrast detectability. The measured organ doses were also used to calculate a reference male effective dose to allow further comparison of the two machines and imaging protocols. The head, chest, and pelvis scans yielded effective doses of 0.04, 7.15, and 3.73 mSv for the XVI, and 0.12, 1.82, and 4.34 mSv for the OBI, respectively.
PACS number: 87.57.uq
Details
- Title: Subtitle
- An organ and effective dose study of XVI and OBI cone‐beam CT systems
- Creators
- Daniel E Hyer - University of FloridaChristopher F Serago - Mayo ClinicSiyong Kim - Mayo ClinicJonathan G Li - University of FloridaDavid E Hintenlang - University of Florida
- Resource Type
- Journal article
- Publication Details
- Journal of applied clinical medical physics, Vol.11(2), pp.181-197
- DOI
- 10.1120/jacmp.v11i2.3183
- PMID
- 20592702
- PMCID
- PMC5719945
- NLM abbreviation
- J Appl Clin Med Phys
- ISSN
- 1526-9914
- eISSN
- 1526-9914
- Publisher
- John Wiley and Sons Inc; Hoboken
- Grant note
- CDC Task 81, TKC 30‐07 185‐01 / Center for Disease Control / TKC Integration Services #DE‐GF07‐05ID14700 / U.S. Department of Energy
- Language
- English
- Date published
- 2010
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Radiation Oncology
- Record Identifier
- 9984047981402771
Metrics
36 Record Views