Assessment of Small-Diameter Aortic Mechanical Prostheses: Physiological Relevance of the Doppler Gradient, Utility of Flow Augmentation, and Limitations of Orifice Area Estimation
Circulation (New York, N.Y.), Vol.98(9), pp.866-872
09/1998
DOI: 10.1161/01.CIR.98.9.866
PMID: 9738641
Abstract
Background
—Noninvasive assessment of functionally stenotic small-diameter aortic mechanical prostheses is complicated by theoretical constraints relating to the hemodynamic relevance of Doppler-derived transprosthetic gradients. To establish the utility of Doppler echocardiography for evaluation of these valves, 20-mm Medtronic Hall and 19-mm St Jude prostheses were studied in vitro and in vivo.
Methods and Results
—Relations between the orifice transprosthetic gradient (equivalent to Doppler), the downstream gradient in the zone of recovered pressure (equivalent to catheter), and fluid mechanical energy losses were examined in vitro. Pressure-flow relations across the 2 prostheses were evaluated by Doppler echocardiography in vivo. For both types of prosthesis in vitro, the orifice was higher than the downstream gradient (
P
<0.001), and fluid mechanical energy losses were as strongly correlated with orifice as with downstream pressure gradients (
r
2
=0.99 for both). Orifice and downstream gradients were higher and fluid mechanical energy losses were larger for the St Jude than the Medtronic Hall valve (all
P
<0.001). Whereas estimated effective orifice areas for the 2 valves in vivo were not significantly different, model-independent dynamic analysis of pressure-flow relations revealed higher gradients for the St Jude than the Medtronic Hall valve at a given flow rate (
P
<0.05).
Conclusions
—Even in the presence of significant pressure recovery, the Doppler-derived gradient across small-diameter aortic mechanical prostheses does have hemodynamic relevance insofar as it reflects myocardial energy expenditure. Small differences in function between stenotic aortic mechanical prostheses, undetectable by conventional orifice area estimations, can be identified by dynamic Doppler echocardiographic analysis of pressure-flow relations.
Details
- Title: Subtitle
- Assessment of Small-Diameter Aortic Mechanical Prostheses: Physiological Relevance of the Doppler Gradient, Utility of Flow Augmentation, and Limitations of Orifice Area Estimation
- Creators
- Richard H. Marcus - University of Chicago Medical CenterRussell S. HeinrichJames BednarzStephen LupovitchJoseph AbruzzoRaphael BorokByron VandenbergRichard E. KerberWilliam PiccioneAjit P. YoganathanRoberto M. Lang
- Resource Type
- Journal article
- Publication Details
- Circulation (New York, N.Y.), Vol.98(9), pp.866-872
- DOI
- 10.1161/01.CIR.98.9.866
- PMID
- 9738641
- ISSN
- 0009-7322
- eISSN
- 1524-4539
- Language
- English
- Date published
- 09/1998
- Academic Unit
- Radiology; Cardiovascular Medicine; Internal Medicine
- Record Identifier
- 9984360156302771
Metrics
19 Record Views