Journal article
Changes in ventilation during use of heat and humidity exchangers
Minerva anestesiologica, Vol.57(1-2), pp.13-16
01/1991
PMID: 2057084
Abstract
The Authors studied in 10 patients during mechanical ventilation the effects of the application of heat and moisture exchanger (HME) in the presence of constant TV and frequency. They observed an increase of PaCO2 and PE, CO2 during the use of HME; this is due to Vd/Vt increase. The Authors conclude that it is mandatory to increase the total ventilation when using HME to avoid dangerous levels of hypercapnia.
Details
- Title: Subtitle
- Changes in ventilation during use of heat and humidity exchangers
- Creators
- M Croci - Ospedale MaggioreF CorradoE SibillaA TibyG VercesiD ProiettiA Vannucci
- Resource Type
- Journal article
- Publication Details
- Minerva anestesiologica, Vol.57(1-2), pp.13-16
- PMID
- 2057084
- ISSN
- 0375-9393
- eISSN
- 1827-1596
- Alternative title
- Variazioni della ventilazione dovute all'uso di scambiatori di umidità e calore
- Language
- Italian
- Date published
- 01/1991
- Academic Unit
- Anesthesia
- Record Identifier
- 9984656607302771
Metrics
10 Record Views