Journal article
Characterizing the Impact of Off-Axis Scan Acquisition on the Reproducibility of Total Retinal Thickness Measurements in SDOCT Volumes
Translational Vision Science & Technology, Vol.4(4), 3
2015
DOI: 10.1167/tvst.4.4.3
PMCID: PMC4525795
PMID: 26257998
Abstract
<p>PURPOSE: Off-axis acquisition of spectral domain optical coherence tomography (SDOCT) images has been shown to increase total retinal thickness (TRT) measurements. We analyzed the reproducibility of TRT measurements obtained using either the retinal pigment epithelium (RPE) or Bruch's membrane as reference surfaces in off-axis scans intentionally acquired through multiple pupil positions.</p>
<p>METHODS: Five volumetric SDOCT scans of the macula were obtained from one eye of 25 normal subjects. One scan was acquired through a central pupil position, while subsequent scans were acquired through four peripheral pupil positions. The internal limiting membrane, the RPE, and Bruch's membrane were segmented using automated approaches. These volumes were registered to each other and the TRT was evaluated in 9 Early Treatment of Diabetic Retinopathy Study (ETDRS) regions. The reproducibility of the TRT obtained using the RPE was computed using the mean difference, coefficient of variation (CV), and the intraclass correlation coefficient (ICC), and compared to those obtained using Bruch's membrane as the reference surface. A secondary set of 1545 SDOCT scans was also analyzed in order to gauge the incidence of off-axis scans in a typical acquisition environment.</p>
<p>RESULTS: The photoreceptor tips were dimmer in off-axis images, which affected the RPE segmentation. The overall mean TRT difference and CV obtained using the RPE were 7.04 ± 4.31 μm and 1.46%, respectively, whereas Bruch's membrane was 1.16 ± 1.00 μm and 0.32%, respectively. The ICCs at the subfoveal TRT were 0.982 and 0.999, respectively. Forty-one percent of the scans in the secondary set showed large tilts (> 6%).</p>
<p>CONCLUSIONS: RPE segmentation is confounded by its proximity to the interdigitation zone, a structure strongly affected by the optical Stiles-Crawford effect. Bruch's membrane, however, is unaffected leading to a more robust segmentation that is less dependent upon pupil position.</p>
<p>TRANSLATIONAL RELEVANCE: The way in which OCT images are acquired can independently affect the accuracy of automated retinal thickness measurements. Assessment of scan angle in a clinical dataset demonstrates that off-axis scans are common, which emphasizes the need for caution when relying on automated thickness parameters when this component of scan acquisition is not controlled for.</p>
Details
- Title: Subtitle
- Characterizing the Impact of Off-Axis Scan Acquisition on the Reproducibility of Total Retinal Thickness Measurements in SDOCT Volumes
- Creators
- Bhavna Antony - University of California, BerkeleyPaul StetsonMichael AbramoffKyungmoo LeeJohanna ColijnGabriëlle BuitendijkCaroline KlaverAustin RoordaBrandon Lujan
- Resource Type
- Journal article
- Publication Details
- Translational Vision Science & Technology, Vol.4(4), 3
- DOI
- 10.1167/tvst.4.4.3
- PMID
- 26257998
- PMCID
- PMC4525795
- NLM abbreviation
- Transl Vis Sci Technol
- ISSN
- 2164-2591
- eISSN
- 2164-2591
- Copyright
- 2015 The Association for Research in Vision and Ophthalmology, Inc.
- Language
- English
- Date published
- 2015
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Electrical and Computer Engineering; Ophthalmology and Visual Sciences
- Record Identifier
- 9983806258402771
Metrics
20 Record Views