Introduction The Medtronic SelectSecure (TM) (Minneapolis, MN, USA) pacing lead (SS) has theoretical advantages compared to conventional (C) transvenous pacing leads (PLs). The study purpose was to determine whether differences in electrical function and lead survival exist between these PLs in a large data set of pediatric and congenital patients. Methods A multicenter historical longitudinal cohort study was performed comparing SS and CPL performance over a 72-month follow-up (FU). Ten centers provided data for both SS and CPL, matched for age, implanted pacing chamber, time period of implantation, and presence of heart disease. Results The cohort consisted of 141 subjects in each group. No statistical differences were observed in age, gender, presence of heart disease, or pacing indication. Atrial and ventricular capture thresholds were stable throughout FU and higher in the SS group (atrial: 0.75 +/- 0.02 vs 0.5 +/- 0.04 V, ventricular: 1.0 +/- 0.04 vs 0.75 +/- 0.04 V), P < 0.001. Group PL sensing thresholds did not differ. The SS group required greater energy to pace (atrial: 0.57 +/- 0.05 vs 0.32 +/- 0.02 mJ, ventricular: 0.83 +/- 0.05 vs 0.56 +/- 0.06 mJ), P = 0.001. Early lead dislodgement and phrenic nerve stimulation were greater in the SS group (P = 0.03). Long-term lead survival was high and similar between the two groups, P = 0.35. Conclusions Long-term survival of both PL was high with a low fracture rate. The SS had excellent electrical function but did show higher capture thresholds and increased energy to pace; these differences are offset by other advantages of the SS PL.
Details
- Title: Subtitle
- Comparison of the Medtronic SelectSecure and conventional pacing leads: Long-term follow-up in a multicenter pediatric and congenital cohort
- Creators
- Jeffrey P. Moak - Children's NationalIan H. Law - University of IowaMartin J. LaPage - University of MichiganFrank Fish - Vanderbilt UniversityIra Shatty - Advocate Children's HospitalAnne M. Dubin - Stanford UniversityAkash Patel - University of California, San FranciscoPeter Fishbach - Division of Cardiology, Sibley Heart Center, Atlanta, Georgia.Nicole Cain - Medical University of South CarolinaChristopher Johnsrude - University of LouisvilleCharles I. Berul - Children's NationalAminata Bangoura - Children's NationalSridhar Hanumanthaiah - Children's NationalRobert McCarter - Children's National
- Resource Type
- Journal article
- Publication Details
- Pacing and clinical electrophysiology, Vol.42(3), pp.356-365
- DOI
- 10.1111/pace.13614
- PMID
- 30680764
- NLM abbreviation
- Pacing Clin Electrophysiol
- ISSN
- 0147-8389
- eISSN
- 1540-8159
- Publisher
- Wiley
- Number of pages
- 10
- Grant note
- Medtronic, Minneapolis, MN; Medtronic
- Language
- English
- Date published
- 03/01/2019
- Academic Unit
- Cardiology; Stead Family Department of Pediatrics
- Record Identifier
- 9984354046402771
Metrics
16 Record Views