Journal article
Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load
Journal of clinical microbiology, Vol.42(6), pp.2774-2776
06/2004
DOI: 10.1128/JCM.42.6.2774-2776.2004
PMCID: PMC427852
PMID: 15184467
Abstract
We compared the performance of two UltraSensitive AMPLICOR HIV-1 MONITOR kits (version 1.5 [v1.5] versus v1.0) by retesting 404 plasma samples with low viral loads (<3,000 copies/ml) with both kits. With 292 samples that initially had <50 copies/ml by the v1.0 kit, the v1.5 assay was more sensitive than the v1.0 assay for samples with human immunodeficiency virus type 1 RNA near the 50-copy/ml cutoff (P = 0.0146). Median numbers of copies per milliliter were similar for 112 samples with 50 to 3,000 copies/ml with no difference in sensitivity with a 200-copy/ml cutoff.
Details
- Title: Subtitle
- Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load
- Creators
- J Brooks Jackson - Johns Hopkins Medical Institutions, Carnegie Building, Room 415, 600 North Wolfe St., Baltimore, MD 21287, USA. bjackso@jhmi.eduEstelle Piwowar-ManningLeTanya Johnson-LewisRoland BassettLisa M DemeterDonald Brambilla
- Resource Type
- Journal article
- Publication Details
- Journal of clinical microbiology, Vol.42(6), pp.2774-2776
- DOI
- 10.1128/JCM.42.6.2774-2776.2004
- PMID
- 15184467
- PMCID
- PMC427852
- NLM abbreviation
- J Clin Microbiol
- ISSN
- 0095-1137
- eISSN
- 1098-660X
- Publisher
- United States
- Language
- English
- Date published
- 06/2004
- Academic Unit
- Pathology; VPMA - Administration
- Record Identifier
- 9984046833002771
Metrics
24 Record Views