Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load
Journal article   Open access   Peer reviewed

Comparison of versions 1.0 AND 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load

J Brooks Jackson, Estelle Piwowar-Manning, LeTanya Johnson-Lewis, Roland Bassett, Lisa M Demeter and Donald Brambilla
Journal of clinical microbiology, Vol.42(6), pp.2774-2776
06/2004
DOI: 10.1128/JCM.42.6.2774-2776.2004
PMCID: PMC427852
PMID: 15184467
url
https://doi.org/10.1128/JCM.42.6.2774-2776.2004View
Published (Version of record) Open Access

Abstract

We compared the performance of two UltraSensitive AMPLICOR HIV-1 MONITOR kits (version 1.5 [v1.5] versus v1.0) by retesting 404 plasma samples with low viral loads (<3,000 copies/ml) with both kits. With 292 samples that initially had <50 copies/ml by the v1.0 kit, the v1.5 assay was more sensitive than the v1.0 assay for samples with human immunodeficiency virus type 1 RNA near the 50-copy/ml cutoff (P = 0.0146). Median numbers of copies per milliliter were similar for 112 samples with 50 to 3,000 copies/ml with no difference in sensitivity with a 200-copy/ml cutoff.
HIV-1 - isolation & purification Humans Reagent Kits, Diagnostic Sensitivity and Specificity RNA, Viral - analysis HIV-1 - genetics Viral Load