Logo image
Demonstration of motion-compensated volumetric modulated arc radiotherapy on an MR-linac
Journal article   Open access   Peer reviewed

Demonstration of motion-compensated volumetric modulated arc radiotherapy on an MR-linac

Pim T.S. Borman, Prescilla Uijtewaal, Jeffrey Snyder, Bryan Allen, Caiden K. Atienza, Peter Woodhead, Daniel E. Hyer, Bas W. Raaymakers and Martin F. Fast
Physics and imaging in radiation oncology, Vol.33, 100729
01/2025
DOI: 10.1016/j.phro.2025.100729
PMCID: PMC11880728
PMID: 40046572
url
https://doi.org/10.1016/j.phro.2025.100729View
Published (Version of record) Open Access

Abstract

Intensity-modulated radiotherapy (IMRT) in combination with magnetic resonance imaging (MRI)-guided gated delivery represents the latest development in the treatment of abdominothoracic tumours on MR-linac. In contrast, volumetric-modulated arc therapy (VMAT) is typically used on conventional linacs due to its superior delivery efficiency and speed. Non-inferior VMAT plans were created in a research treatment planning system for eight lung cancer patients previously treated on an MR-linac. VMAT plans were delivered on a moving dosimeter using respiratory multi-leaf collimator (MLC) tracking. VMAT with MLC tracking achieved an average 2%/2 mm local gamma pass rate of 93% relative to planned dose with a delivery efficiency of 83%.
Lung Cancer MLC tracking MR-linac VMAT