Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system
Medical physics (Lancaster), Vol.42(12), pp.7022-7033
12/01/2015
DOI: 10.1118/1.4935534
PMCID: PMC4644157
PMID: 26632057
Abstract
Purpose: Image registration between standard x-ray fluoroscopy and transesophageal echocardiography (TEE) has recently been proposed. Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system designed for cardiac procedures. This study presents a method for 3D registration of SBDX and TEE images based on the tomosynthesis and 3D tracking capabilities of SBDX.
Methods: The registration algorithm utilizes the stack of tomosynthetic planes produced by the SBDX system to estimate the physical 3D coordinates of salient key-points on the TEE probe. The key-points are used to arrive at an initial estimate of the probe pose, which is then refined using a 2D/3D registration method adapted for inverse geometry fluoroscopy. A phantom study was conducted to evaluate probe pose estimation accuracy relative to the ground truth, as defined by a set of coregistered fiducial markers. This experiment was conducted with varying probe poses and levels of signal difference-to-noise ratio (SDNR). Additional phantom and in vivo studies were performed to evaluate the correspondence of catheter tip positions in TEE and x-ray images following registration of the two modalities.
Results: Target registration error (TRE) was used to characterize both pose estimation and registration accuracy. In the study of pose estimation accuracy, successful pose estimates (3D TRE < 5.0 mm) were obtained in 97% of cases when the SDNR was 5.9 or higher in seven out of eight poses. Under these conditions, 3D TRE was 2.32 +/- 1.88 mm, and 2D (projection) TRE was 1.61 +/- 1.36 mm. Probe localization error along the source-detector axis was 0.87 +/- 1.31 mm. For the in vivo experiments, mean 3D TRE ranged from 2.6 to 4.6 mm and mean 2D TRE ranged from 1.1 to 1.6 mm. Anatomy extracted from the echo images appeared well aligned when projected onto the SBDX images.
Conclusions: Full 6 DOF image registration between SBDX and TEE is feasible and accurate to within 5 mm. Future studies will focus on real-time implementation and application-specific analysis. (C) 2015 American Association of Physicists in Medicine.
Details
- Title: Subtitle
- Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system
- Creators
- Charles R. Hatt - University of Wisconsin–MadisonMichael T. Tomkowiak - University of Wisconsin–MadisonDavid A. P. Dunkerley - Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USAJordan M. Slagowski - University of Wisconsin–MadisonTobias Funk - Triple Ring TechnologiesAmish N. Raval - University of Wisconsin–MadisonMichael A. Speidel - University of Wisconsin–Madison
- Resource Type
- Journal article
- Publication Details
- Medical physics (Lancaster), Vol.42(12), pp.7022-7033
- DOI
- 10.1118/1.4935534
- PMID
- 26632057
- PMCID
- PMC4644157
- NLM abbreviation
- Med Phys
- ISSN
- 0094-2405
- eISSN
- 2473-4209
- Publisher
- AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
- Number of pages
- 12
- Grant note
- R01 HL084022 / NIH; United States Department of Health & Human Services; National Institutes of Health (NIH) - USA R01HL084022 / NATIONAL HEART, LUNG, AND BLOOD INSTITUTE; United States Department of Health & Human Services; National Institutes of Health (NIH) - USA; NIH National Heart Lung & Blood Institute (NHLBI)
- Language
- English
- Date published
- 12/01/2015
- Academic Unit
- Radiation Oncology
- Record Identifier
- 9984312962402771
Metrics
9 Record Views