Journal article
Dosimetric aspects of a rotating beam splitter used in tangential field breast treatment
Medical physics (Lancaster), Vol.14(5), pp.879-883
09/1987
DOI: 10.1118/1.596017
PMID: 3683321
Abstract
A rotating beam splitter was designed and fabricated for use in treating tangential breast fields on an AECL Theratron‐80 cobalt teletherapy unit. Its dosimetric properties were studied using a 0.6‐cm3 Baldwin–Farmer ionization chamber with Keithley electrometer and a Scanditronix RFA‐3 three‐dimensional water phantom scanner with semiconductor detector. An aluminum plate, which held the semicircular rotating 5‐HVL (half‐value layer) lead block, extended to the phantom surface (80‐cm source–surface distance). The beam was blocked directly along the central axis and also at distances up to 7.5 mm off‐axis, corresponding to the projected extent of the 1.5‐cm‐diam source. The penumbra at the central ray and at each off‐axis point was measured at d max and at 5‐cm depth in water. A reduction in the penumbra from 8 to about 2 mm for 20×20 cm2 beam was observed regardless of the off‐axis distance of the block. Isodose distributions obtained for various field sizes indicated that the percent depth doses of the split fields agree well with the equivalent squares of the irradiated field sizes. Output measurements in water and in air indicated that scatter from the aluminum plate more than compensates for the reduction in backscatter factor, due to the decrease in irradiated area when the beam splitter is used. Isodose curves in various planes were obtained at clinically useful rotational angles of the beam splitter. Computer generated isodose curves have been obtained that match the measured curves to be used in treatment planning.
Details
- Title: Subtitle
- Dosimetric aspects of a rotating beam splitter used in tangential field breast treatment
- Creators
- Shirish K Jani - Division of Radiation Oncology, Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, Iowa 52242Edward C Pennington - Division of Radiation Oncology, Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, Iowa 52242Antonio P Vigliotti - Division of Radiation Oncology, Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, Iowa 52242David H Hussey - Division of Radiation Oncology, Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, Iowa 52242
- Resource Type
- Journal article
- Publication Details
- Medical physics (Lancaster), Vol.14(5), pp.879-883
- DOI
- 10.1118/1.596017
- PMID
- 3683321
- ISSN
- 0094-2405
- eISSN
- 2473-4209
- Number of pages
- 5
- Date published
- 09/1987
- Academic Unit
- Radiation Oncology
- Record Identifier
- 9984047857602771
Metrics
14 Record Views