Effectiveness of averaging strategies to reduce variance in retinal nerve fibre layer thickness measurements using spectral-domain optical coherence tomography
Journal article   Peer reviewed

Effectiveness of averaging strategies to reduce variance in retinal nerve fibre layer thickness measurements using spectral-domain optical coherence tomography

Berthold Pemp, Randy H Kardon, Karl Kircher, Elisabeth Pernicka, Ursula Schmidt-Erfurth and Andreas Reitner
Graefe's archive for clinical and experimental ophthalmology, Vol.251(7), pp.1841-1848
07/2013
DOI: 10.1007/s00417-013-2337-0
PMID: 23589277

View Online

Abstract

Automated detection of subtle changes in peripapillary retinal nerve fibre layer thickness (RNFLT) over time using optical coherence tomography (OCT) is limited by inherent image quality before layer segmentation, stabilization of the scan on the peripapillary retina and its precise placement on repeated scans. The present study evaluates image quality and reproducibility of spectral domain (SD)-OCT comparing different rates of automatic real-time tracking (ART). Peripapillary RNFLT was measured in 40 healthy eyes on six different days using SD-OCT with an eye-tracking system. Image brightness of OCT with unaveraged single frame B-scans was compared to images using ART of 16 B-scans and 100 averaged frames. Short-term and day-to-day reproducibility was evaluated by calculation of intraindividual coefficients of variation (CV) and intraclass correlation coefficients (ICC) for single measurements as well as for seven repeated measurements per study day. Image brightness, short-term reproducibility, and day-to-day reproducibility were significantly improved using ART of 100 frames compared to one and 16 frames. Short-term CV was reduced from 0.94 ± 0.31 % and 0.91 ± 0.54 % in scans of one and 16 frames to 0.56 ± 0.42 % in scans of 100 averaged frames (P ≤ 0.003 each). Day-to-day CV was reduced from 0.98 ± 0.86 % and 0.78 ± 0.56 % to 0.53 ± 0.43 % (P ≤ 0.022 each). The range of ICC was 0.94 to 0.99. Sample size calculations for detecting changes of RNFLT over time in the range of 2 to 5 μm were performed based on intraindividual variability. Image quality and reproducibility of mean peripapillary RNFLT measurements using SD-OCT is improved by averaging OCT images with eye-tracking compared to unaveraged single frame images. Further improvement is achieved by increasing the amount of frames per measurement, and by averaging values of repeated measurements per session. These strategies may allow a more accurate evaluation of RNFLT reduction in clinical trials observing optic nerve degeneration.
Retinal Ganglion Cells - cytology Optic Disk - anatomy & histology Reproducibility of Results Humans Adult Female Male Reference Values Tomography, Optical Coherence - statistics & numerical data Nerve Fibers Visual Acuity Observer Variation