Logo image
Hemolytic anemia after operation for aortic dissection using Teflon felt strips
Journal article   Peer reviewed

Hemolytic anemia after operation for aortic dissection using Teflon felt strips

Yuki Nakamura, Hitoshi Ogino, Hitoshi Matsuda, Kenji Minatoya, Hiroaki Sasaki and Soichiro Kitamura
The Annals of thoracic surgery, Vol.85(5), pp.1784-1787
05/01/2008
DOI: 10.1016/j.athoracsur.2007.11.057
PMID: 18442587

View Online

Abstract

We report three cases of hemolytic anemia caused by anastomotic stenosis after surgical treatment for aortic dissection in which internal and external Teflon (DuPont, Wilmington, DE) felt strips were used for reinforcement of the aortic stump. To detect this complication, laboratory findings typical of red cell fragmentation syndrome as well as appropriate imaging modalities are necessary. As a precaution, it is necessary to be meticulous when stitching the internal felt strip. (c) 2008 by The Society of Thoracic Surgeons.
Cardiac & Cardiovascular Systems Cardiovascular System & Cardiology Life Sciences & Biomedicine Respiratory System Science & Technology Surgery