Abstract
We introduce an intradural approach to spinal cord stimulation for the relief of intractable pain, and describe the biophysical rationale that underlies its design and performance requirements. The proposed device relies on wireless, inductive coupling between a pial surface implant and its epidural controller, and we present the results of benchtop experiments that demonstrate the ability to transmit and receive a frequency-modulated 1.6 MHz carrier signal between micro-coil antennae scaled to the ≈ 1 cm dimensions of the implant, at power levels of about 5 mW. Plans for materials selection, microfabrication, and other aspects of future development are presented and discussed.
Details
- Title: Subtitle
- Intradural approach to selective stimulation in the spinal cord for treatment of intractable pain: design principles and wireless protocol
- Creators
- M Howard - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAM Utz - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAT Brennan - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAB Dalm - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAS Viljoen - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAN Jeffery - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USAG Gillies - University of Iowa Hospitals and Clinics, Department of Neurosurgery, 200 Hawkins Drive, Iowa City, Iowa 52242-1086, USA
- Resource Type
- Journal article
- Publication Details
- Journal of applied physics, Vol.110(4), pp.044702-044702-12
- Publisher
- American Institute of Physics
- DOI
- 10.1063/1.3626469
- ISSN
- 0021-8979
- eISSN
- 1089-7550
- Date published
- 08/30/2011
- Academic Unit
- Neurosurgery; Iowa Neuroscience Institute; Anesthesia; Neurology; Otolaryngology
- Record Identifier
- 9984070226502771
Metrics
6 Record Views