Journal article
Kinematic comparison of walking on uneven ground using powered and unpowered prostheses
Clinical biomechanics (Bristol), Vol.28(4), pp.467-472
04/2013
DOI: 10.1016/j.clinbiomech.2013.03.005
PMID: 23602128
Abstract
Recent research has focused on the design of intelligent prosthetic ankle devices with the goal of adapting behavior of the device to accommodate all walking surfaces that an individual encounters in daily life. To date, no studies have looked at how such devices perform on uneven terrain.
11 young adults with unilateral transtibial amputation participated in two data collection sessions spaced approximately 3 weeks apart. In each session they walked across a loose rock surface at three controlled speeds. In the first session, they wore a passive, energy storage and return prosthesis and in the second, they wore a powered prosthesis (BiOM, iWalk, Bedford, MA, USA).
Subjects had a 10% faster self-selected walking speed when wearing the powered (1.16 m/s) compared to unpowered prosthesis (1.05 m/s; p=0.031). They walked with increased ankle plantarflexion on their prosthetic limb throughout the gait cycle when wearing the powered compared to unpowered prosthesis. This was especially evident in the increased plantarflexion during push-off (p<0.001). There was a small (<3°), but statistically significant decrease in knee flexion during early stance when wearing the powered device (p=0.045). Otherwise, the kinematics of the knee and hip were nearly identical when wearing the different devices. Subjects had decreased medial-lateral motion of their center of mass when wearing the powered prosthesis (p=0.020), but there were no differences in medial-lateral margins of stability between the devices (p=0.662).
Subjects did not significantly alter their proximal joint kinematics on this irregular surface as a result of the addition of power.
Details
- Title: Subtitle
- Kinematic comparison of walking on uneven ground using powered and unpowered prostheses
- Creators
- Deanna H Gates - Department of Orthopaedics and Rehabilitation, Brooke Army Medical Center, Ft. Sam Houston, TX 78234, USAJennifer M AldridgeJason M Wilken
- Resource Type
- Journal article
- Publication Details
- Clinical biomechanics (Bristol), Vol.28(4), pp.467-472
- Publisher
- England
- DOI
- 10.1016/j.clinbiomech.2013.03.005
- PMID
- 23602128
- ISSN
- 0268-0033
- eISSN
- 1879-1271
- Grant note
- DOI: 10.13039/100006965, name: U.S. Army Telemedicine & Advanced Technology Research Center; name: Center for Rehabilitation Sciences Research
- Language
- English
- Date published
- 04/2013
- Academic Unit
- Physical Therapy and Rehabilitation Science
- Record Identifier
- 9984047754302771
Metrics
15 Record Views