Limit of Detection and Threshold for Positivity of the Centers for Disease Control and Prevention Assay for Factor VIII Inhibitors
Journal of thrombosis and haemostasis, Vol.15(10), pp.1971-1976
09/14/2017
DOI: 10.1111/jth.13795
PMCID: PMC5716470
PMID: 28795528
Abstract
Essentials Immunologic methods detect factor VIII (FVIII) antibodies in some inhibitor-negative specimens. Specimens were tested by modified Nijmegen-Bethesda assay (NBA) and fluorescence immunoassay. The NBA with preanalytical heat inactivation detects FVIII inhibitors down to 0.2 NBU. IgG4 frequency validates the established threshold for positivity of ≥ 0.5 NBU for this NBA. Summary: Background The Bethesda assay for measurement of factor VIII inhibitors called for quantification of positive inhibitors by using dilutions producing 25–75% residual activity (RA), corresponding to 0.4–2.0 Bethesda units, with the use of ‘more sensitive methods’ for samples with RA closer to 100% being recommended. The Nijmegen modification (Nijmegen-Bethesda assay [NBA]) changed the reagents used but not these calculations. Some specimens negative by the NBA have been shown to have FVIII antibodies detectable with sensitive immunologic methods. Objective To examine the performance at very low inhibitor titers of the Centers for Disease Control and Prevention (CDC)-modified NBA (CDC-NBA), which includes preanalytic heat inactivation to liberate bound anti-FVIII antibodies. Methods Specimens with known inhibitors were tested with the CDC-NBA. IgG4 anti-FVIII antibodies were measured by fluorescence immunoassay (FLI). Results Diluted inhibitors showed linearity below 0.4 Nijmegen-Bethesda units (NBU). With four statistical methods, the limit of detection of the CDC-NBA was determined to be 0.2 NBU. IgG4 anti-FVIII antibodies, which correlate most strongly with functional inhibitors, were present at rates above the background rate of healthy controls in specimens with titers ≥ 0.2 NBU and showed an increase in frequency from 14.3% at 0.4 NBU to 67% at the established threshold for positivity of 0.5 NBU. Conclusions The CDC-NBA can detect inhibitors down to 0.2 NBU. The FLI, which is more sensitive, demonstrates anti-FVIII IgG4 in some patients with negative (< 0.5) NBU. The sharp increase in IgG4 frequency between 0.4 and 0.5 NBU validates the established threshold for positivity of ≥ 0.5 NBU for the CDC-NBA, supporting the need for method-specific thresholds.
Details
- Title: Subtitle
- Limit of Detection and Threshold for Positivity of the Centers for Disease Control and Prevention Assay for Factor VIII Inhibitors
- Creators
- Connie H. Miller - National Center on Birth Defects and Developmental DisabilitiesBrian Boylan - National Center on Birth Defects and Developmental DisabilitiesAmy D. Shapiro - Indiana Hemophilia and Thrombosis CenterSteven R. Lentz - University of IowaBrian M. Wicklund - Kansas City Regional Hemophilia Center, Kansas City, MO, USA.Hemophilia Inhibitor Research Study Investigators
- Resource Type
- Journal article
- Publication Details
- Journal of thrombosis and haemostasis, Vol.15(10), pp.1971-1976
- DOI
- 10.1111/jth.13795
- PMID
- 28795528
- PMCID
- PMC5716470
- NLM abbreviation
- J Thromb Haemost
- ISSN
- 1538-7933
- eISSN
- 1538-7836
- Grant note
- name: CDC Foundation; DOI: 10.13039/100004319, name: Pfizer; DOI: 10.13039/100007658, name: Baxter Healthcare Corporation
- Language
- English
- Date published
- 09/14/2017
- Academic Unit
- Hematology, Oncology, and Blood & Marrow Transplantation; Internal Medicine
- Record Identifier
- 9984359920502771
Metrics
11 Record Views