Abstract
Minimally traumatic surgical techniques and advances in cochlear implant (CI) electrode array designs have allowed acoustic hearing present in a CI candidate prior to surgery to be preserved postoperatively. As a result, these patients benefit from combined electric-acoustic stimulation (EAS) postoperatively. However, 30% to 40% of EAS CI users experience a partial loss of hearing up to 30 dB after surgery. This additional hearing loss is generally not severe enough to preclude use of acoustic amplification; however, it can still impact EAS benefits. The use of electrocochleography (ECoG) measures of peripheral hair cell and neural auditory function have shed insight into the pathophysiology of postimplant loss of residual acoustic hearing. The present study aims to assess the long-term stability of ECoG measures and to establish ECoG as an objective method of monitoring residual hearing over the course of EAS CI use. We hypothesize that repeated measures of ECoG should remain stable over time for EAS CI users with stable postoperative hearing preservation. We also hypothesize that changes in behavioral audiometry for EAS CI users with loss of residual hearing should also be reflected in changes in ECoG measures.
A pool of 40 subjects implanted under hearing preservation protocol was included in the study. Subjects were seen at postoperative visits for behavioral audiometry and ECoG recordings. Test sessions occurred 0.5, 1, 3, 6, 12 months, and annually after 12 months postoperatively. Changes in pure-tone behavioral audiometric thresholds relative to baseline were used to classify subjects into two groups: one group with stable acoustic hearing and another group with loss of acoustic hearing. At each test session, ECoG amplitude growth functions for several low-frequency stimuli were obtained. The threshold, slope, and suprathreshold amplitude at a fixed stimulation level was obtained from each growth function at each time point. Longitudinal linear mixed effects models were used to study trends in ECoG thresholds, slopes, and amplitudes for subjects with stable hearing and subjects with hearing loss.
Preoperative, behavioral audiometry indicated that subjects had an average low-frequency pure-tone average (125 to 500 Hz) of 40.88 ± 13.12 dB HL. Postoperatively, results showed that ECoG thresholds and amplitudes were stable in EAS CI users with preserved residual hearing. ECoG thresholds increased (worsened) while ECoG amplitudes decreased (worsened) for those with delayed hearing loss. The slope did not distinguish between EAS CI users with stable hearing and subjects with delayed loss of hearing.
These results provide a new application of postoperative ECoG as an objective tool to monitor residual hearing and understand the pathophysiology of delayed hearing loss. While our measures were conducted with custom-designed in-house equipment, CI companies are also designing and implementing hardware and software adaptations to conduct ECoG recordings. Thus, postoperative ECoG recordings can potentially be integrated into clinical practice.
Details
- Title: Subtitle
- Longitudinal Electrocochleography as an Objective Measure of Serial Behavioral Audiometry in Electro-Acoustic Stimulation Patients
- Creators
- Viral D Tejani - University of IowaJeong-Seo Kim - University of Iowa Hospitals and ClinicsChristine P Etler - University of IowaJeffrey Skidmore - The Ohio State University Wexner Medical CenterYi Yuan - The Ohio State University Wexner Medical CenterShuman He - The Ohio State University Wexner Medical CenterMarlan R Hansen - University of IowaBruce J Gantz - University of IowaPaul J Abbas - University of IowaCarolyn J Brown - University of Iowa
- Resource Type
- Journal article
- Publication Details
- Ear and hearing, Vol.44(5), pp.1014-1028
- DOI
- 10.1097/AUD.0000000000001342
- PMID
- 36790447
- PMCID
- PMC10425573
- NLM abbreviation
- Ear Hear
- eISSN
- 1538-4667
- Language
- English
- Electronic publication date
- 02/15/2023
- Date published
- 09/2023
- Academic Unit
- Communication Sciences and Disorders; Molecular Physiology and Biophysics; Neurosurgery; Otolaryngology
- Record Identifier
- 9984367560202771
Metrics
11 Record Views