MR safety and imaging of Neuroform stents at 3T
Journal article   Peer reviewed

MR safety and imaging of Neuroform stents at 3T

Arvind NEHRA, Christopher J MORAN, Dewitte T CROSS and Colin P DERDEYN
American journal of neuroradiology : AJNR, Vol.25(9), pp.1476-1478
2004
PMID: 15502124

View Online

Abstract

The Neuroform stent is a self-expanding nitinol stent designed for use in wide-necked intracranial aneurysms. Heating and imaging artifacts were evaluated by using a porcine carotid artery aneurysm model in a 3T MR system. A suspended Neuroform stent was tested for deflection. No heating was measured, and no evidence of deflection of the stent was found. Imaging artifacts were minimal. MR imaging in patients with Neuroform stent-treated aneurysms is safe and feasible.
Vision Nervous System Fundamental and applied biological sciences. Psychology Radiodiagnosis. Nmr imagery. Nmr spectrometry Investigative techniques, diagnostic techniques (general aspects) Psychology. Psychoanalysis. Psychiatry Psychology. Psychophysiology Biological and medical sciences Perception Medical sciences