Performance characteristics of two double-walled infant incubators
Critical care medicine, Vol.11(8), pp.663-667
08/1983
DOI: 10.1097/00003246-198308000-00017
PMID: 6409507
Abstract
Two double-walled incubators, the Air Shields C-100 and the Ohio IC, were evaluated for performance characteristics. The Ohio incubator heated more rapidly from room temperature to 36°C, but overshot the preset air temperature and produced greater fluctuation in air temperature due to the operation of the servocontrol system at equilibrium. Neither incubator produced excessive air currents, sound levels, or carbon dioxide accumulation.