Journal article
Production of 16 N for instrument calibration
Radiochimica acta, Vol.92(4-6), pp.229-232
04/01/2004
DOI: 10.1524/ract.92.4.229.35583
Abstract
Abstract
A convenient source of -7 second [
16
N]-N
2
is described, employing the deuteron irradiation of a flow-through nitrogen target, suitable for the calibration of gamma detectors to 7 MeV and beta spectrometers to 10 MeV. Its possible presence as a contaminant in
15
O-gases bears watching.
Details
- Title: Subtitle
- Production of 16 N for instrument calibration
- Creators
- R. J. NicklesT. E. BarnhartM. A. Avila-RodriguesA. K. ConverseR. SundaresanJ. A. NyeD. W. DickA. D. Roberts
- Resource Type
- Journal article
- Publication Details
- Radiochimica acta, Vol.92(4-6), pp.229-232
- DOI
- 10.1524/ract.92.4.229.35583
- ISSN
- 0033-8230
- eISSN
- 2193-3405
- Language
- English
- Date published
- 04/01/2004
- Academic Unit
- Radiology; Pharmaceutical Sciences and Experimental Therapeutics
- Record Identifier
- 9984318709602771
Metrics
6 Record Views