Radionuclide calibrator measurements of 18F in a 3 ml plastic syringe
Applied radiation and isotopes, Vol.66(6), pp.988-993
2008
DOI: 10.1016/j.apradiso.2008.02.046
PMID: 18375134
Abstract
This paper discusses the calibration of Capintec radionuclide calibrators for the measurement of
18F in the pharmacy and clinic. In support of a planned regional comparison, a secondary laboratory was set up at Oak Ridge National Laboratories in Oak Ridge, TN. The laboratory was used to prepare 1
mL
18F-fluorodeoxyglucose sources in 3
mL plastic BD syringes (Becton, Dickinson and Company, Franklin Lakes, New Jersey, USA) and to determine dial settings in three models of radionuclide calibrators. Measurements were verified by comparing radionuclide calibrator measurements of a 5
mL ampoule source with previous results. The CIEMAT/NIST method of tritium efficiency tracing was used for activity determinations.
Details
- Title: Subtitle
- Radionuclide calibrator measurements of 18F in a 3 ml plastic syringe
- Creators
- J.T Cessna - National Institute of Standards and Technology, 100 Bureau Drive MS8462, Gaithersburg, MD 20899, USAM.K Schultz - National Institute of Standards and Technology, 100 Bureau Drive MS8462, Gaithersburg, MD 20899, USAT Leslie - PETNET Solutions, University of Tennessee Medical Center at Knoxville, Knoxville, TN 37920, USAN Bores - Oak Ridge National Laboratories, P.O. Box 2008 MS6366, Oak Ridge, TN 37831, USA
- Resource Type
- Journal article
- Publication Details
- Applied radiation and isotopes, Vol.66(6), pp.988-993
- Publisher
- Elsevier Ltd
- DOI
- 10.1016/j.apradiso.2008.02.046
- PMID
- 18375134
- ISSN
- 0969-8043
- eISSN
- 1872-9800
- Language
- English
- Date published
- 2008
- Academic Unit
- Radiology; Stead Family Department of Pediatrics; Radiation Oncology
- Record Identifier
- 9984046907102771
Metrics
16 Record Views