Reliability of Spinopelvic Measurements That May Influence the Cup Position in Total Hip Arthroplasty
The Journal of arthroplasty, Vol.35(12), pp.3758-3764
12/2020
DOI: 10.1016/j.arth.2020.06.056
PMID: 32888750
Abstract
Spinopelvic pathology has been identified as a potential risk factor for instability after total hip arthroplasty. Spinopelvic radiographic parameters used to diagnose spinopelvic disease may also impact optimal cup placement. The purpose of this study was to assess the arthroplasty surgeon accuracy in making spinopelvic measurements.
Five fellowship-trained or in-training arthroplasty surgeons reviewed 44 lateral lumbar radiographs in two sessions. All evaluators were instructed how to perform measurements but had little experience in doing so. Traditional measurements included the pelvic tilt (PT), pelvic incidence (PI), and sacral slope (SS), and novel measurements included the acetabular ante-inclination (AI) and pelvic femoral angle. Surgeon measurements were compared with those made by experienced engineers. intraclass coefficients (ICCs) were calculated and interpreted.
The interobserver reliability for 4 of 5 of the spinopelvic parameters was good to excellent (ICC >0.75) with the highest reliability seen for PI measurement (ICC = 0.939). Only moderate interobserver reliability was observed for AI measurement (ICC = 0.559). Intraobserver reliability was the highest for the PI and SS, ranging from moderate to good (ICC, 0.718 to 0.896). The lowest intraobserver reliability was seen for the AI (ICC range, 0.026 to 0.545) and pelvic femoral angle (ICC range, 0.035 to 0.828). Surgeon measurements of PT and SS were compared with engineer measurements with extremely poor correlation observed (ICC <0 for all).
Surgeon intraobserver and interobserver reliability in making novel radiographic measurements was low compared with traditional radiographic measurements. Surgeon reliability in making traditional measurements of the PT and SS was very poor compared with experienced assessors using software-based measurements.
Details
- Title: Subtitle
- Reliability of Spinopelvic Measurements That May Influence the Cup Position in Total Hip Arthroplasty
- Creators
- Lindsay T. Kleeman-Forsthuber - Colorado Joint ReplacementJacob M. Elkins - University of IowaTodd M. Miner - Colorado Joint ReplacementCharlie C. Yang - Colorado Joint ReplacementJason M. Jennings - Colorado Joint ReplacementDouglas A. Dennis - University of Colorado Anschutz Medical Campus
- Resource Type
- Journal article
- Publication Details
- The Journal of arthroplasty, Vol.35(12), pp.3758-3764
- Publisher
- Elsevier Inc
- DOI
- 10.1016/j.arth.2020.06.056
- PMID
- 32888750
- ISSN
- 0883-5403
- eISSN
- 1532-8406
- Language
- English
- Date published
- 12/2020
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Orthopedics and Rehabilitation
- Record Identifier
- 9984304712902771
Metrics
9 Record Views