Journal article
Spatial Variation of Resolution and Noise in Multi–Detector Row Spiral CT
Academic radiology, Vol.10(6), pp.607-613
2003
DOI: 10.1016/S1076-6332(03)80078-8
PMID: 12809413
Abstract
The authors performed this study to evaluate an approach for measuring the variations of three-dimensional spatial resolution and image noise throughout a field of view imaged with multi–detector row spiral computed tomographic (CT) scanners.
The authors designed a phantom (diameter, 320 mm) that contained 37 metallic spheres (diameter, approximately 0.8 mm) positioned between two disks made of a material with attenuation being that of water. One sphere was located at the isocenter of the phantom, and the rest were evenly spaced in three concentric rings with diameters of 100, 200, and 300 mm, respectively. The phantom was imaged with two widely used multi–detector row CT scanners by using a standard protocol and four variations of that protocol. Because a recently developed theory holds that image resolution should be proportional to the square root of the trace of the covariance matrix of a point spread function, the authors developed a software package to segment high-attenuation spheres from the CT image volume and compute point spread functions from blurred images of the spheres. Three-dimensional spatial resolution and image noise were calculated as a function of radial distance within the field of view.
Resolution and noise were quantified in the resultant CT image volumes and found to be nonisotropic, with worse resolution and less noise occurring at the periphery of the field of view.
The method enabled measurement of variations in spatial resolution and of their distribution on images obtained with multi–detector row CT scanners. These findings may contribute to the development of an improved algorithm for image reconstruction.
Details
- Title: Subtitle
- Spatial Variation of Resolution and Noise in Multi–Detector Row Spiral CT
- Creators
- John F Meinel - Department of Biomedical Engineering, University of Iowa College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242, USAGe Wang - Department of Biomedical Engineering, University of Iowa College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242, USAMing Jiang - School of Mathematics, Peking University, Beijing, ChinaTroy Frei - Department of Radiology, University of Iowa College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242, USAMichael Vannier - Department of Radiology, University of Iowa College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242, USAEric Hoffman - Department of Biomedical Engineering, University of Iowa College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242, USA
- Resource Type
- Journal article
- Publication Details
- Academic radiology, Vol.10(6), pp.607-613
- Publisher
- Elsevier Inc
- DOI
- 10.1016/S1076-6332(03)80078-8
- PMID
- 12809413
- ISSN
- 1076-6332
- eISSN
- 1878-4046
- Language
- English
- Date published
- 2003
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Radiology; Internal Medicine
- Record Identifier
- 9984051992002771
Metrics
34 Record Views