Journal article
Stress distribution in cylindrical and conical implants under rotational micromovement with different boundary conditions and bone properties: 3-D FEA
Computer methods in biomechanics and biomedical engineering, Vol.20(8), pp.893-900
06/11/2017
DOI: 10.1080/10255842.2017.1309394
PMID: 28347170
Abstract
Factors related to micromovements at bone-implant interface have been studied because they are considered adverse to osseointegration. Simplifications are commonly observed in these FEA evaluations. The aim of this study was to clarify the influence of FEA parameters (boundary conditions and bone properties) on the stress distribution in peri-implant bone tissue when micromovements are simulated in implants with different geometries. Three-dimensional models of an anterior section of the jaw with cylindrical or conical titanium implants (4.1mm in width and 11mm in length) were created. Micromovement (50, 150, or 250m) was applied to the implant. The FEA parameters studied were linear vs. non-linear analyses, isotropic vs. orthogonal anisotropic bone, friction coefficient (0.3) vs. frictionless bone-implant contact. Data from von Mises, shear, maximum, and minimum principal stresses in the peri-implant bone tissue were compared. Linear analyses presented a relevant increase of the stress values, regardless of the bone properties. Frictionless contact reduced the stress values in non-linear analysis. Isotropic bone presented lower stress than orthogonal anisotropic. Conical implants behave better, in regard to compressive stresses (minimum principal), than cylindrical ones, except for nonlinear analyses when micromovement of 150 and 250m were simulated. The stress values raised as the micromovement amplitude increased. Non-linear analysis, presence of frictional contact and orthogonal anisotropic bone, evaluated through maximum and minimum principal stress should be used as FEA parameters for implant-micromovement studies.
Details
- Title: Subtitle
- Stress distribution in cylindrical and conical implants under rotational micromovement with different boundary conditions and bone properties: 3-D FEA
- Creators
- Mateus Bertolini Fernandes dos Santos - Univ Fed Pelotas, Sch Dent, Pelotas, BrazilGabriel de Oliveira Meloto - Univ Estadual Campinas, Piracicaba Dent Sch, Piracicaba, BrazilAtais Bacchi - Atitus EducaçãoLourenco Correr-Sobrinho - Universidade Estadual de Campinas (UNICAMP)
- Resource Type
- Journal article
- Publication Details
- Computer methods in biomechanics and biomedical engineering, Vol.20(8), pp.893-900
- DOI
- 10.1080/10255842.2017.1309394
- PMID
- 28347170
- NLM abbreviation
- Comput Methods Biomech Biomed Engin
- ISSN
- 1025-5842
- eISSN
- 1476-8259
- Publisher
- Taylor & Francis
- Number of pages
- 8
- Grant note
- 2013/08031-2; 2012/19374-5 / FAPESP (Sao Paulo Research Foundation); Fundacao de Amparo a Pesquisa do Estado de Sao Paulo (FAPESP) 12/19374-5 / Fundacao de Amparo a Pesquisa do Estado de Sao Paulo (FAPESP)
- Language
- English
- Date published
- 06/11/2017
- Academic Unit
- Prosthodontics
- Record Identifier
- 9984946625902771
Metrics
6 Record Views