Test-retest variability of blue-on-yellow perimetry is greater than white-on-white perimetry in normal subjects
Journal article   Peer reviewed

Test-retest variability of blue-on-yellow perimetry is greater than white-on-white perimetry in normal subjects

Young H Kwon, Hyun J Park, Aliza Jap, Seyda Ugurlu and Joseph Caprioli
American journal of ophthalmology, Vol.126(1), pp.29-36
07/1998
DOI: 10.1016/S0002-9394(98)00062-2
PMID: 9683146

View Online

Abstract

To compare long-term fluctuation of blue-on-yellow automated perimetry with white-on-white automated perimetry in normal subjects. White-on-white and blue-on-yellow automated perimetry were performed on a Humphrey Visual Field Analyzer and an Octopus perimeter, both modified for blue-on-yellow perimetry. The study sample consisted of 31 eyes of 31 normal subjects for the Humphrey perimeter and 33 eyes of 33 normal subjects for the Octopus perimeter. After one practice session, each subject completed four testing sessions over a period of 2 to 8 weeks, each separated by at least 1 day. Each testing session consisted of both white-on-white and blue-on-yellow perimetry performed on one eye; the order of the tests was alternated for successive sessions. Long-term fluctuation (expressed as statistical variance) was calculated for each test location. Intersubject variability (expressed as statistical variance) across all subjects was determined for each test location. On the Humphrey perimeter, the long term fluctuation for blue-on-yellow perimetry (4.07 +/- 3.07 dB2) was significantly greater than that for white-on-white perimetry (1.97 +/- 0.99 dB2; P < .001). Long-term fluctuation increased as a function of eccentricity for both blue-on-yellow and white-on-white perimetry. Short-term fluctuation was significantly greater for blue-on-yellow (0.46 +/- 0.25 dB) than that for white-on-white perimetry (0.29 +/- 0.19 dB; P < .02). Finally, the intersubject variability was significantly greater in blue-on-yellow (13.2 +/- 2.8 dB2) than it was in white-on-white perimetry (4.25 +/- 1.13 dB2; P < .001). Similar results were found with the Octopus perimeter. Long-term fluctuation and short-term fluctuation of blue-on-yellow perimetry are greater than those of white-on-white perimetry in normal subjects. The increased long-term fluctuation requires appropriate statistical approaches when evaluating serial change of blue-on-yellow perimetry.
Visual Field Tests - methods Visual Fields - physiology Reproducibility of Results Prospective Studies Humans Middle Aged Sensitivity and Specificity Adult Female Male Reference Values Visual Field Tests - instrumentation

Details

Metrics

Logo image