The effective dynamic ranges of standard automated perimetry sizes III and V and motion and matrix perimetry
Journal article   Open access   Peer reviewed

The effective dynamic ranges of standard automated perimetry sizes III and V and motion and matrix perimetry

Michael Wall, Kimberly R Woodward, Carrie K Doyle and Gideon Zamba
Archives of ophthalmology (1960), Vol.128(5), pp.570-576
05/2010
DOI: 10.1001/archophthalmol.2010.71
PMID: 20457977
url
https://doi.org/10.1001/archophthalmol.2010.71View
Published (Version of record) Open Access

Abstract

To establish the associations between threshold estimates of 4 perimetric tests and to define and compare the tests' effective dynamic ranges. We examined 152 patients with glaucoma and 80 controls using standard automated perimetry (SAP) with stimulus size III, SAP with size V, and motion and matrix perimetry. We explored the intertest associations using principal-components analysis. We defined the effective dynamic range bottom using the frequency of 0-dB trials on retest. We defined the upper effective dynamic range as a value above which fewer than 0.5% of the values fall in the controls. We also calculated the number of discriminable steps from normal to the floor value of the perimeter. The association between SAP III and V was approximately linear up to a sensitivity of about 20 dB on both tests and with motion and matrix perimetry up to about 25 dB from 0 dB. While the upper bounds were similar among the tests, size V SAP had a lower floor and more discriminable steps. The effective dynamic range of SAP III is substantially less than its physically tested limits. Size V stimuli have a greater effective dynamic range than size III by about 1 log unit and have about twice as many discriminable steps.
Motion Visual Field Tests - methods Sensory Thresholds Humans Middle Aged Male Glaucoma - diagnosis Vision Disorders - diagnosis Vision Disorders - physiopathology Visual Fields Sensitivity and Specificity Aged, 80 and over Adult Female Aged Visual Field Tests - instrumentation Glaucoma - physiopathology

Details

Metrics

Logo image