Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T
bioRxiv
Cold Spring Harbor Laboratory
09/22/2026
DOI: 10.64898/2026.09.21.753210
PMCID: PMC13622145
PMID: 42817996
Abstract
Subject motion and gradient heating-induced frequency and phase offsets result in spectral misalignment during single-voxel MR spectroscopy (SVS) acquisitions. Several methods have been presented to align the spectra, but their performance on a 7T system is unclear. This study aimed to evaluate the practical importance of four retrospective correction methods, namely creatine fitting, residual water, spectral registration, and cross-correlation. SVS data were collected from 127 participants (88/39 female/male, 39±15 years) using a semi-localization by adiabatic selective refocusing sequence at 7T. Changes in the spectral linewidth, signal-to-noise ratio (SNR), similarity (mean similarity matrix (SI
)), and metabolite quantification (concentration, relative Cramer-Rao lower bounds (rCRLB)) after correction were evaluated. A Wilcoxon signed rank test was used to evaluate these changes. The
-values were adjusted using the false discovery rate (
<0.05) for multiple comparisons. For significant changes, the effect size was further calculated using the Rosenthal formula. The results demonstrated that all correction methods resulted in a significant decrease in the spectral linewidth and significant increases in the spectral SNR and SI
. The mean changes by using the four methods were 0.76-0.99 Hz for the spectral linewidth change (effect sizes:0.73-0.86), 5.79-6.95 for the SNR change (effect size:0.60-0.72), and 0.03-0.03 for the SI
change (effect size:0.87-0.87). The correction also significantly increased the estimates of the metabolite concentrations of PCr, Gln, Glu, GPC, Ins, NAA, Scyllo, and Tau (effect size:0.25-0.57), and significantly decreased the rCRLBs (effect size:0.26-0.42) of Scyllo, Tau, and Gly. These results indicate that each method provides practical benefits in improving the spectral linewidth, SNR, and similarity by showing a strong to very strong effect of the improvements (effect size>0.60). The correction only showed a weak to moderate effect of the changes in the metabolite quantification (effect size<0.60), suggesting that whether the correction results in more reliable metabolite quantification requires further validation.
Details
- Title: Subtitle
- Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T
- Creators
- Chu-Yu Lee - University of IowaJia Xu - University of Iowa, RadiologyBaolian YangRalph NoeskeVincent A Magnotta - University of Iowa
- Resource Type
- Preprint
- Publication Details
- bioRxiv
- DOI
- 10.64898/2026.09.21.753210
- PMID
- 42817996
- PMCID
- PMC13622145
- NLM abbreviation
- bioRxiv
- ISSN
- 2692-8205
- eISSN
- 2692-8205
- Publisher
- Cold Spring Harbor Laboratory; United States
- Language
- English
- Date posted
- 09/22/2026
- Academic Unit
- Roy J. Carver Department of Biomedical Engineering; Radiology; Psychiatry; Iowa Neuroscience Institute
- Record Identifier
- 9985240901902771
Metrics
1 Record Views